Fields of application
The dust concentration measuring device PFM 97 W is used for the continuous measurement of dust contents, especially for the emission measurement according to TI air, 13th, 17th and 27th BImSchV as well as for the dust concentration measurement in process monitorings. Parallel to the dust measurement flow and temperature of the smoke gas are registered.
temperature: 0 … 300 °C, velocity: 0 … 30 m/s flow: 0 … 1.000.000 m³/h, dust i.O.: 0 … 15 (max. 500) mg/m³, dust i.n.: 0 … 15/45/150/500 mg/m³
dust: redundant measurement with 2 triboelectric probes; flow: differential pressure; temperature: Pt 100
‘-20 … +50 °C
5 x 4 ... 20 mA, (including 2 x dust, temperature, flow) , galvanically isolated
6 potential-free contacts (failure, maintenance, limit value 1 and 2 / maintenance request, measuring range)
110 VAC, 230 VAC or 24 VDC
TI air, 13th, 17th and 27th BImSchV